technical proportions

All loadcells are individually tested and are supplied with a certificate of calibration.

example:        
model     1.00.058    
capacity     4T    
serial No.   1234    
output     16.6736 mV/V  
zero balance   <+/_ 1.00% of full scale
creep      <+/_ 0.05% of full scale
non linearity   <+/_ 2.50% of full scale
hysteresis   <+/_ 1.20% of full scale
repeatability   <+/_ 0.03% of full scale
temperature effect on output < 50 PPM/°C of applied load
temperature effect on zero   < 50 PPM/°C of applied load
temperature range: operating -20 °C to +60 °C  
temperature range: compensated -10 °C to +50 °C  
safe overload   120% of capacity  
imput impendance   750 +/- 10 Ohms  
output impendance   700 +/- 5 Ohms  
insulation impedance   ≥ 5000 M Ohms  
exitation: reccomended   10 V DC/AC  
exitation: maximum   20 V DC/AC  
           
Colour code:        
red + exitation        
black - exitation        
green + signal        
white - signal